Vacuum Test Chambers

The DHT® Vacuum Test Chambers are advanced environmental simulation systems specifically engineered to replicate high-altitude, low-pressure conditions while delivering precise temperature and humidity control. Designed with a focus on long-term operational stability, minimal maintenance requirements, and superior performance, these chambers are ideal for rigorous testing applications across aerospace, defense, electronics, and export-oriented manufacturing sectors.
Featuring a broad internal volume range from 100 to 1500 liters, the chambers accommodate a wide variety of test specimens and configurations. They operate across a wide temperature range of -70°C to +180°C, with customizable starting points available at -20°C, -40°C, or 0°C to suit specific testing protocols. The system ensures excellent temperature uniformity—within ±1.0°C—even under dynamic vacuum conditions, allowing accurate, repeatable measurements of component behavior under simulated altitude stress.
Engineered with high-efficiency vacuum systems, intelligent control algorithms, and robust stainless steel interiors, the DHT® Vacuum Test Chambers deliver reliable performance for critical tasks such as avionics validation, sensor calibration, battery insulation assessment, and altitude adaptability verification of high-reliability products.

0/5 (0 Reviews)

Benefits

Advanced Control System: Equipped with an intelligent controller imported from Europe, the chamber is easy to operate, user-friendly, and highly stable, ensuring precise and controlled data during the testing process.

Optimized Air Circulation System: The specially designed air circulation system provides uniform airflow and precise testing conditions for the samples, ensuring the accuracy and repeatability of test results.

Precise Humidity Control: The optimized wet-bulb and dry-bulb humidity measurement system ensures highly accurate humidity control, essential for reliable testing under low-pressure conditions.

Air-Cooled Refrigeration System: Featuring a high-efficiency air-cooled refrigeration system, the chamber maintains stable operating temperatures even in low-pressure test environments, enhancing the durability of the equipment.

Self-Diagnosis System: Integrated with a self-diagnosis system, the chamber can detect and alert users to any operational issues, facilitating maintenance and repair, and ensuring continuous testing operations.

Features Video

Specifications

Temperature Range (°C) 0~180;-40~180;
-65~180;-70~180
Temperature Fluctuation (°C) ≤±1.0 (atmospheric pressure, no load)
Temperature Uniformity (°C) ±2.0 (atmospheric pressure, no load)
Temperature Gradient (°C) ≤2.0
Temperature Rate (°C/min) Heating/Cooling Rate: 0.5 ~ 1.0°C/min
Non-linear Adjustable (atmospheric pressure, no load)
Temperature Deviation (°C) ≤2.0 (atmospheric pressure, no load)
During low pressure composite testing: ±3
Pressure Recovery Rate (kPa/min) ≤10
Pressure Range (kPa) Customizable
Pressure Accuracy (kPa) 0.1
Pressure Variation Rate (kPa/min) ≤10
Available Temperature Range for Humidity Control (°C) 15 ~ 95
Relative Humidity Range (%RH) 20~98
Humidity Accuracy (%RH) ±1.0 ~±3.0